KLEINWACHTER CPM374 CHARGED PLATE MONITOR 
with integrated Electro-Fieldmeter 
SYSTEM PROF. DR. ING. HANS KLEINW?CHTER 
 
The CHARGE PLATE Monitor 374 measures the voltage potential on its integrated Plate-Electrode according the Fieldmill-Influence-Principal and is suitable to determine the Static Decay Time from a selected higher voltage threshold to a lower voltage threshold level. 
  
CPM374 
Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H
 
The CHARGE PLATE Monitor CPM374 consists of an integrated timer, a high-voltage unit and an electro-fieldmill-sensor and achieves therefore the high impedance of 10
15 W to measure electrostatic voltage potentials and static decay times of air ionizer equipment. 
 
The unit meets the requirements of the Standard IEC 61340-5-1 and the American Standard EOS/ESD S.3.1.  
 
For the vertical measurement use the included three Distance pieces. Screw them on the backside of the unit. 
 
Important: The high voltage unit is protected by a high ohmic current limiting resistor and therefore there is no danger when touching the charged plate. 
 
The CPM374 is battery powered and due to its compact design, it is very suitable for the use in industrial areas. It can also be operated at a power supply (230/115 VAC). The CPM374 is menu-driven and will be controlled by means of 2 switches for simple operation. All selected parameters will be shown in the 2 rows LCD-display for a better orientation of the operator.  
 
The enclosed software enables the remote control operation of the CPM374 and data processing by means of a PC. 
 
2.1  Switching on  
The CHARGE PLATE Monitor CPM374 is equipped with a NiCd-battery and ready for operation - use the enclosed power supply only to charge the battery. The display background illumination is not operating when battery powered. The battery operation features an automatic switch-off function, if the switches were not touched for 
approximay 10 minutes. when the unit is run on the power supply (230/115 VAC) the background illumination is switched on while the automatic switch-off function is disabled. 
 
Press button B to switch on the CPM374. 
  
1.  Description 
2.  Operation 
2.2  Decay Time and Offset Voltage 
 
The software version will be shown shortly before the measurement mode and the command to initiate the measurement appear on the display: 
 
DECAY TIME: AUTO 
PRESS START! 
 
The initially setting is saved in the EEPROM and will be shown after each switch-on. 
 
If the auto-mode is selected, as indicated in the example above, then the time will be measured for the positive voltage decay first and then followed automatically for the negative. This mode is recommended for checking air ionizer equipment, as the Startbutton has to be pushed on time only to perform a complete check of ionizers including the determination of the Offset Voltage. 
 
Offset-Voltage is the voltage potential, which remains after a certain time on the plate electrode. It is an indication for the balance of positive and negative ions. If the Offset- Voltage is positive, then an excess of positive ions emanate from the ionizer equipment - the other way round for negative. Please consider the specification of the ionizer for a qualified evaluation. 
 
The Start-button has to be pressed in order to charge the plate-electrode by means of the integrated high-voltage unit. The timer setting and the applied voltage potential are shown on the display: 
 
TIMER: 1000-100V 
U = XXXXV 
 
The timer starts as soon as the upper threshold value (1000 volt in this case) is passed: 
 
TIMER=XX.Xs 
U = XXXXV 
 
and will stop the decay time when reaching the lower value (100 volt). The voltage setting for the timer could be changed in the set-up function. 
 
The offset-voltage will be measured after a certain lapse time at the end of the static decay time measurement: 
OFFSET=XXXs 
U = XXXXV 
 
The measurement series is now finished and all relevant parameters can be obtained from the display, e.g. to qualify an air ionizer equipment. The offset-voltage UOFF and the belonging lapse time will be finally displayed in the first row for 10 seconds or until a new measurement was initiated by pressing button A. 
 
 The decay time appears in the second row as POS DECAY or as NEG DECAY or both  alternating every second, if the auto-mode was pre-selected: 
 
UOFF=XXXV T=XXXs 
POS.DECAY =XX.Xs 
 
2.3  Saving measurement values  
The measured values can be saved in the internal memory, therefore the request will follow after each measurement accordingly:  
SAVE VALUES? 
A<YES NO>B 
 
press the button A to start the saving. In case that the memory is full (99 entries), the display will indicate for approximay 2 seconds: 
 
MEMORY FULL! 
 
All files have to be deleted in order to perform the saving of the last measurement: The following request will appear on the screen:  
DELETE FILES? 
A<YES NO>B 
 
In case, that the full memory should not be deleted, please press B and the unit will be ready to take the next measurement - the last measurement results will disappear and cannot be saved. 
 
Important: It is recommended to transfer all files to your PC after each measurement interval by means of the enclosed software. Thus will allow to delete files of 
the CPM374 at the next measurement interval without loosing important data. 
 
Press A to delete the full memory of the CPM374 and the proposal for the file number of the last measurement will appear: 
 
FILE NO: XX 
A<OK CHANGE>B 
 
Press A to confirm this file number or change with each press of B until you reach the right file number. Finally, press A to confirm the right file number. 
In order to reach faster a certain file number press B, then the ten digit can be changed by pressing A while the single digit can be changed with B. when you have reached the right file number, wait for approximay 3 seconds until the display change back as stated above and confirm the setting by pressing A. 
 
Press A and B simultaneously to start the set-up function. The CPM374 will ask for performing the set-up function:  
SET-UP ? 
A<YES NO>B  
 
3.1  SET-UP  
If the set-up function was started, the last Static Decay and Offset setting will appear (in this case Mode = Auto), which can be confirmed again by pressing A:  
MODE = AUTO 
A<OK CHANGE>B 
 
If B will be pressed, the following options are shown one after the other: 
MODE = POS 
MODE = NEG 
MODE = VOLTMETER 
MODE = AUTO 
 
The decay time for only positive respectively negative potentials with the following offset voltage are measured in the MODE = POS and MODE = NEG. In the MODE = 
VOLTMETER, the CPM374 will measure permanently the voltage potential on the plate electrode - Use a measurement lead connected to the plate electrode to operate the 
CPM374 as an Electrometer (High Impedance Voltmeter), e.g. monitoring the change of voltage potential on human bodies, caused by movements. 
 
If you have selected the decay time measurement (MODE = POS, MODE = NEG or MODE = AUTO), the starting point for the voltage potential, from where on the decay time 
should be measured, will be asked first:  
START: U=1000V 
A<OK CHANGE>B 
 
Press A to confirm the proposed starting point. If B will be pressed, the following options are shown one after the other:  
500V - 550V - ... - 900V - 950V - 1000V - 500V 
 
and the selected value has to be confirmed by pressing A . 
 
 
3.  SET-UP Function  
The same procedure will follow selecting the stop point, that means at which potential the decay time should be stopped:  
STOP: U=100V 
A<OK CHANGE>B 
 
Press A to confirm the proposed stop point. If B will be pressed, the following options are shown one after the other:  
0V - 50V - 100V - ... -400V - 450V - 500V - 0V 
 
and the selected value has to be confirmed by pressing A .  
Finally, the CPM 374 will ask for the lapse time to determine the offset voltage:  
OFFSET: XXXs 
A<OK CHANGE>B 
  
Press A to confirm the proposed lapse time. If B will be pressed, the following options are shown one after the other:  
10s - 20s - ... - 50s - 1min - 2min- ... - 10min - 10s 
 
and the selected value has to be confirmed by pressing A  . 
The chosen parameters for Start, Stop and Offset will be shown for approximay 3 sec, 
e.g.: 
TIMER: 1000-100V 
OFFSET: XXXs 
 
and finally the unit will recheck if all parameters are chosen correctly:  
TIMER+OFFSET OK? 
A<YES NO>B 
 
The selected parameters will be confirmed by pressing A and could be changed again with B. 
When A was pressed, the unit will ask for clearing the memory, as the stored data files are based on the previous setting:  
CLEAR MEMORY? 
A<YES NO>B 
 
 
If B was pressed, the previous setting will remain the same, the memory was not deleted and the new parameter change is ignored, the following message will appear: 
 PARAMETER 
NOT CHANGED ! 
 
If A was pressed, then following message will advice that the parameter change was taken over and the memory is cleared accordingly, the following message will appear for approximay 2 sec:  
PARAMETER 
CHANGED ! 
 
The display will show now the new decay time setting and the request to press start! 
 
3.2  Calibration  
If NO>B was chosen for the set-up function, the unit will ask for the Electro-Fieldmeter calibration:  
EFM CALIBRATION? 
A<YES NO>B 
 
Press B if a calibration is not required. Press A to initiate the automatic offset calibration, the following message will appear for approximay 1 sec:  
PLEASE WAIT 
CALIBRATION 
 
Important: The plate electrode must be free of foreign voltage connection during calibration. 
 
The display will show the new decay time setting and the request to press start! 
 
If the electrofield-meter is defected, the following message will appear:  
EFM DEFECT! 
REPAIR IS NEEDED 
 
If the power of the NiCd Battery is less than 6,7 volts, the unit will provide the following message alternating every second with the actual first row: 
 
LOW BATTERY!  
 
4.  Other messages on the display 
The battery has to be recharged, but the ongoing or the upcoming measurement could be finished first.  
 
Important:  
Please use the enclosed battery charger only for charging the batteries. The batteries will also be charged when the CPM374 is switched off. The max. charging time is 14 hours, but the unit could be longer connected to the battery charger as it consists of a charging control device. It is recommended to use the CPM374 until LOW BATTERY! appears on screen. This will maintain the max. battery capacitance over a longer period. 
 
If the battery voltage will become less than 6,1 volts, the following message will appear on screen for approximay 2 seconds:  
AUTO OFF 
LOW BATTERY! 
 
and the unit will switch off automatically to prevent a total discharge of the battery. A total discharge can cause damage or destroy the battery! 
Connect the unit by means of the provided cable to a COM socket of a PC. Insert the floppy disk and select the right drive (e.g. A:). Type ?START“ and press ENTER, the software will be installed on your PC. 
 
Requirements: 
min. 486 processor, 3.5“ floppy disk drive, 2 MB RAM, DOS or Windows. the program is self explaining. 
We provide 12 month warranty from date of receiving the unit, if the unit was handled properly. The damage of the NiCd-battery caused by improper handling as well as 
mechanical damage on the unit are excluded from warranty. Warranty is not provided, if the housing was opened. 
 
5.  PC-Operating 
6.  Warranty 
The unit has to be grounded sufficiently for accurate measurements of the electrical field and its polarity. Therefore the unit has to be connected to ground potential by means of its grounding socket. 
 
Housing :  Aluminum-housing 152mm x 152mm x 152mm (L x W x H) 
Weight :  1,5 kg 
Display :  Alphanumeric Display, 2 rows with 16 digits each, 
display area 100 mm x 24 mm 
Measurement Range :  Voltage potentials up to ± 1200 V, accuracy ± 2,5% 
PC-Interface :  Serial port, 9 pol Sub D – socket 
Battery :  7,2 V / 850 mAh 
last approximay 4 hours for continuous operation when 
battery was fully charged, charging time max. 14 hours 
Battery Charger :  12 V / 500 mA 
Load Current:  max. 350mA, max. 150 mA for battery operating 
Operating Temperature:  0 ... 40°C 
Relative Humidity:  0 ... 60% 
 
·  CHARGE PLATE Monitor CPM 374 
·  Power transformer prim: 230VAC sec: 12VDC 500mA 
·  Teflon insulated measurement lead, 1m 
·  Crocodile clip 
·  Grounding cable 
·  Serial PC-port cable 
·  3 Spacers for horizontal decay time measurement 
·  Operating Instructions 
·  Certificate of factory calibration 
 
  
7.  Grounding 
8.  Specification 
9.  Parts of Delivery 
 
The measurement unit is not approved for the use in explosive areas! 
The unit is not approved for the use in power utilities! 
This unit is not suitable to measure alternating fields with a frequency higher than 1 Hz. 
Sparking discharge to the electro-fieldmill-sensor has to be avoided. 
 
The Teflon-supports of the plate electrode must be maintained clean, therefore do not touch with fingers. All Teflon spacers and the teflonisolated wire to the plate have to be cleaned carefully with Acetone or similar solvents in case that the voltage on the charged plate drops in a rate of greater than 100V / minute under normal circumstances (no external lead connection to the plate electrode and no presence of air ionizers). The interval for factory calibration is normally 1 year. 
 
 
Please note the following advice when checking air ionizers: 
The presence of charged objects as well as personnel could cause failure on the measurement results caused by influence of voltage. The unit should be therefore always 
grounded to avoid this phenomena from the housing of the unit itself. 
 
 
10.  Safety Advice 
11.  Maintenance 
12.  Important Notice 
 
KONFORMIT?TSERKL?RUNG 
 
Die Firma 
 
Kleinw?chter GmbH 
Krummattstr. 9 
79688 Hausen i.W. 
 
erkl?rt in alleiniger Verantwortung, dass das Produkt 
 
Charge Plate Monitor CPM 374 
Messger?te zur Messung des elektrischen Gleichspannungsfeldes 
 
auf die sich diese Erkl?rung bezieht, die Forderung 
 
der EWG-Richtlinie über die elektromagnetische Vertr?glichkeit 
Richtlinie des Rates vom 3.5.1989 (89/336) Stand: Juli 1993 
 
erfüllt, insbesondere der Normen 
 
EN 61010 ?Safety“ 
EN 55011 Gruppe 1, Klasse B 
EN 50082-2/92 
 
für ISM-Ger?te. 
 
Die Firma Kleinw?chter GmbH h?lt folgende technischen Dokumentationen zur Einsicht bereit 
 
·  vorschriftsm??ige Bedienungsanleitung 
·  Baupl?ne 
·  technische Dokumentationen 
 
Hausen, den 09.01.2002 
 
 
Jürgen Brunner 
techn. Gesch?ftsführer 
 
 
KLEINWACHTER
Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H


中國電子行業儀器儀表供應商——上海堅融實業有限公司      :www.atfcdc.cn  
凌志堅 產品     :        
:jet1718  地址:上海市閔行區閔北工業園區紀王鎮紀西周涇 9 號/68 號 

上一篇:德國Kleinwachter TOM-600靜電電阻測試儀手冊說明書

下一篇:DESCO50671/50672離子風機操作說明書手冊

產品分類

聯系我們

在線客服:

主站蜘蛛池模板: 恒温恒湿试验箱_高低温试验箱_恒温恒湿箱-东莞市高天试验设备有限公司 | 复合肥,化肥厂,复合肥批发,化肥代理,复合肥品牌-红四方 | 滤芯,过滤器,滤油机,贺德克滤芯,精密滤芯_新乡市宇清流体净化技术有限公司 | 热处理温控箱,热处理控制箱厂家-吴江市兴达电热设备厂 | PSI渗透压仪,TPS酸度计,美国CHAI PCR仪,渗透压仪厂家_价格,微生物快速检测仪-华泰和合(北京)商贸有限公司 | 硬度计_影像测量仪_维氏硬度计_佛山市精测计量仪器设备有限公司厂家 | 不锈钢轴流风机,不锈钢电机-许昌光维防爆电机有限公司(原许昌光维特种电机技术有限公司) | 深圳3D打印服务-3D打印加工-手板模型加工厂-悟空打印坊 | 智慧水务|智慧供排水利信息化|水厂软硬件系统-上海敢创 | 搪玻璃冷凝器_厂家-越宏化工设备| 塑料检查井_双扣聚氯乙烯增强管_双壁波纹管-河南中盈塑料制品有限公司 | 工业用品一站式采购平台|南创工品汇-官网|广州南创 | 液压中心架,数控中心架,自定心中心架-烟台恒阳机电设计有限公司 行星搅拌机,双行星搅拌机,动力混合机,无锡米克斯行星搅拌机生产厂家 | 冷凝水循环试验箱-冷凝水试验箱-可编程高低温试验箱厂家-上海巨为(www.juweigroup.com) | 水轮机密封网 | 水轮机密封产品研发生产厂家 | RV减速机-蜗轮蜗杆减速机-洗车机减速机-减速机厂家-艾思捷 | 东风体检车厂家_公共卫生体检车_医院体检车_移动体检车-锦沅科贸 | 单锥双螺旋混合机_双螺旋锥形混合机-无锡新洋设备科技有限公司 | 预制舱-电力集装箱预制舱-模块化预制舱生产厂家-腾达电器设备 | 蒸压釜-陶粒板隔墙板蒸压釜-山东鑫泰鑫智能装备有限公司 | 袋式过滤器,自清洗过滤器,保安过滤器,篮式过滤器,气体过滤器,全自动过滤器,反冲洗过滤器,管道过滤器,无锡驰业环保科技有限公司 | 上海赞永| 联系我们-腾龙公司上分客服微信19116098882 | 河南道路标志牌_交通路标牌_交通标志牌厂家-郑州路畅交通 | 浙江上沪阀门有限公司| 硫化罐-胶管硫化罐-山东鑫泰鑫智能装备有限公司 | 物联网卡_物联网卡购买平台_移动物联网卡办理_移动联通电信流量卡通信模组采购平台? | 六维力传感器_三维力传感器_二维力传感器-南京神源生智能科技有限公司 | 轻型地埋电缆故障测试仪,频响法绕组变形测试仪,静荷式卧式拉力试验机-扬州苏电 | 工业电炉,台车式电炉_厂家-淄博申华工业电炉有限公司 | 深圳市源和塑胶电子有限公司-首页| 找培训机构_找学习课程_励普教育 | 空气能采暖,热泵烘干机,空气源热水机组|设备|厂家,东莞高温热泵_正旭新能源 | 汽车润滑油厂家-机油/润滑油代理-高性能机油-领驰慧润滑科技(河北)有限公司 | 浙江华锤电器有限公司_地磅称重设备_防作弊地磅_浙江地磅售后维修_无人值守扫码过磅系统_浙江源头地磅厂家_浙江工厂直营地磅 | 压片机_高速_单冲_双层_花篮式_多功能旋转压片机-上海天九压片机厂家 | 膜结构_ETFE膜结构_膜结构厂家_膜结构设计-深圳市烨兴智能空间技术有限公司 | 钢绞线万能材料试验机-全自动恒应力两用机-混凝土恒应力压力试验机-北京科达京威科技发展有限公司 | 桨叶搅拌机_螺旋挤压/方盒旋切造粒机厂家-无锡市鸿诚输送机械有限公司 | 冰晶石|碱性嫩黄闪蒸干燥机-有机垃圾烘干设备-草酸钙盘式干燥机-常州市宝康干燥 | 临海涌泉蜜桔官网|涌泉蜜桔微商批发代理|涌泉蜜桔供应链|涌泉蜜桔一件代发 |